Identification of organic pigments in tattoo inks and permanent make-up using laser desorption ionisation mass spectrometry (2024)

Version Changes

Revised. Amendments from Version 1

In response to reviewer comments, the title and some text passages were changed to clarify that LDI- instead of MALDI-TOF-MS was used in most cases. In the abstract and section "market survey" some minor changes regarding the ban of pigments were included.

Abstract

Nowadays, about 12% of the European and 20% of the US population are tattooed. Rising concerns regarding consumer safety, led to legal restrictions on tattoo and permanent make-up (PMU) inks. Restrictions also include bans on certain colourants. Both ink types use organic pigments for colour-giving, plus inorganic pigments for white and black and colour tones. Pigments are only sparingly soluble in common solvents and occur as suspended particles in the ink matrix. Their detection and identification therefore pose a major challenge for laboratories involved in monitoring the legal compliance of tattoo inks and PMU. We overcame this challenge by developing a direct laser desorption ionisation time-of-flight mass spectrometry method, which included an easy sample clean up. The method proved to be capable of detecting and identifying organic pigments in almost all of the tested ink samples. Method validation and routine deployment during market surveys showed the method to be fit for purpose. Pigment screening of 396 tattoo inks and 55 PMU taken from the Swiss market between 2009 and 2017 lead to the following conclusions: Pigment variety is much greater in tattoo inks (18) than in PMU (10); four prohibited pigments (Pigment Green 7, Pigment Red 122, Pigment Violet 19 and 23) were found in both ink types; for PMU, these four pigments made up 12% of the pigment findings, compared to 32% for tattoo inks. Therefore, legal compliance of PMU was at a higher level. A comparison of pigments found with those declared on tattoo ink labels clearly showed that banned pigments are rarely declared, but rather masked by listing non present legal pigments and label forging; therefore, highlighting the urgency of widespread market controls.

Keywords: tattoo inks, PMUs, colour pigments, MALDI-TOF

Introduction

Over the last two decades, tattoos have become increasingly popular among younger people, with estimates stating that 12% of the European and more than 20% of the US population are tattooed1. These figures, the widespread on-line sales of tattoo related products, and the invasive techniques used during tattooing - where inks are injected into the skin’s dermis - have led to serious concerns on the safety of tattoo inks and permanent make-up (PMU). This finally compelled several European nations to initiate specific legal regulations, which came into effect during the last ten years and are all based on the European Resolution (ResAP 2008)2,3. For colourants, namely pigments, being the ingredients, which give tattoo inks and PMU the desired effect, regulations are summarised as follows: a ban on colourants that can form aromatic amines under reductive cleavage, a negative list for specific colourants and a ban for colourants with restricted use in cosmetics. Despite this welcome development, safety awareness in this field is still not adequate compared to cosmetics, which have been subjected to stringent regulations for decades, even though their application forms are solely non-invasive. Regulations for cosmetics also require manufacturers to evaluate the consumer safety of their cosmetic products and ingredients. In contrast to this, none of the ingredients used in tattoo inks have ever been tested to ensure their health safety when injected into the skin. Furthermore, the generalised assumption that pigments as insoluble colourants pose no health risk, does not hold true in the dermis. Pigment fading and transport of organic pigments from the skin to regional lymph nodes have been well documented for tattoos4,5. Studies have reported that diarylide pigments degrade under sunlight forming a variety of products, some of which are known to be toxic or carcinogenic6. This finding has yet to be incorporated into regulations.

Pigments are at best only sparingly soluble in common solvents, which narrow the choice of suitable analytical techniques. So far the identification of tattoo pigments in inks has mainly been carried out using Raman as well as Fourier transform infrared (FT-IR) spectroscopy7,8. Raman spectroscopy has also been used for matching the profile of tattoo inks with dermatome shave biopsies from patients with allergic reactions9. Another recent paper describes the successful use of pyrolysis gas chromatography (py-GC/MS)10. While FT-IR and Raman Spectroscopy are very suitable for the identification of single pigments, their successful use for pigment mixtures, often present in tattoo inks, has yet to be proven. Direct laser desorption or matrix assisted laser desorption ionisation time-of-flight mass spectrometry (LDI-TOF-MS or MALDI-TOF–MS) has been reported for identifying pigments in automotive coatings11 or art work1214. In this article, we present a validated LDI-TOF-MS approach for the identification of pigments in tattoo inks and PMU. Repeated market surveys in Switzerland showed the method to be fit for purpose.

Methods

Chemicals

As reference materials, more than 150 individual pigments from different producers (mainly BASF (Basel, Switzerland), Clariant (Muttenz, Switzerland) supplied by Omya (Oftringen, Switzerland), Kremer Pigmente (Aichstetten, Germany) and Sigma–Aldrich (Buchs, Switzerland)) were collected over the past years. Mass calibration was performed with the following dyes all from Sigma-Aldrich: Crystal Violet (C.I. 42555), Basic Blue 7 (C.I. 42595), Ethyl violet (C.I. 42600), Basic Blue 11 (C.I. 44040), Basic red 2 (C.I. 50240), Basic Blue 17 (C.I. 52040). MALDI matrix 2,5-Dihydroxybenzoic acid (DHB, 99%) was purchased from Sigma–Aldrich. Ethanol (EtOH) and methanol (MeOH), both analytical grades, were from J.T. Baker (Center Valley, USA).

Suspensions and solutions

Stock suspensions and/or solutions of reference pigments in MeOH (1 mg mL-1) were individually prepared. For mass calibration, a mixture containing six pigments at levels of 50 μg mL-1 for C.I. 42555, C.I. 42595 and C.I. 42600, of 250 μg mL-1 for C.I. 50240 and C.I. 52040 and of 100 μg mL-1 for C.I. 44040 was prepared by mixing homogenous aliquots of each stock suspension and diluting with MeOH.

Samples and sample preparation

Tattoo inks and PMU were taken from the Swiss market over the last eight years. An overview of the analysed samples is given inTable 1.

Ink samples were shaken intensely by hand for about one minute. Then, about 25 µL (20–30 mg) of the suspension or 20 mg of PMU were weighed into a tared 2 mL-tube, mixed with 1 mL of EtOH, sonicated (5 min, 25°C) and centrifuged (5 min, 15000 rpm). The supernatant was removed and the residue containing the colour pigments was diluted with 1 mL of MeOH. After vortexing for some seconds, 1 µL duplicates of the homogenous suspension were immediately spotted onto a steel target plate, air dried at room temperature and used for LDI-TOF-MS.

Pigment verification with standard addition

20 µL of the methanolic ink suspension were transferred to a 96-Well reaction plate (MicroAmp Optical, Applied Biosystems) and mixed with 10 – 50 µL of the appropriate reference pigment suspension (10 mg mL-1 MeOH). One microliter of the homogenous suspension was used for LDI-TOF-MS.

Improvement of detection using DHB as a matrix

20 µL of the methanolic ink suspension were transferred to the reaction plate, mixed with the same volume of DHB (10 mg mL-1 MeOH) and 1 µL of the homogenous suspension was used for MALDI-TOF-MS.

MALDI-TOF-MS

Pigment mass spectra were obtained using a MALDI-TOF Mass Spectrometry Axima™ Confidence machine (Shimadzu-Biotech Corp., Kyoto, Japan) with detection in the reflectron mode with pulsed extraction (optimised at 450 Da) at a N2-laser frequency of 50 Hz and within a mass range from 50 to 2000 Da. The transmission of the laser power was in the range of 60 – 100 units (33 – 55%). The target plate was scanned by the laser (diameter of 30 μm) in the regular rectangular mode and serpentine style (1000 × 1000 μm, spacing 166.666 μm and 49 points). A minimum of 20 laser shots was accumulated per profile and for each sample a total of 50 mass profiles was averaged and processed using Launchpad™ version 2.9.3 software (Shimadzu-Biotech Corp., Kyoto, Japan). This software was also used for peak processing of all raw spectra with the following settings: the advanced scenario was chosen from the parent peak clean up menu, peak width was set to 3 channels, smoothing filter width to 2 channels, baseline filter width to 10 channels, and the threshold apex was chosen as the peak detection method. The threshold apex peak detection was set as a dynamic type and the offset was set to 0.300 mV with a response factor of 1.0. The processed spectra were exported as peak lists with m/z values for each peak and signal intensity in the ASCII format. Calibration was conducted for each target plate using spectra of the reference standard mixture consisting of the following exact masses: 271.1137, 315.1609, 372.2439, 422.2596, 456.3378 and 478.3222 amu.

Identification of pigments with NIST library software

Exported ASCII files from MALDI-TOF-MS were converted to NIST software compatible peak lists with a mass resolution of 1 D (*.msp), and the identification of pigments was carried out with our home-made mass spectra library using the software of NIST 2.0 (Standard Reference Data Program of the National Institute of Standards and Technology, USA).

HPLC

More or less soluble colourants detected by LDI-TOF-MS were occasionally confirmed by high performance liquid chromatography (HPLC). Approximately 5 mg of tattoo or PMU samples were extracted with different solvents, starting with N,N-Dimethylformamide, 1-Chloronaphthalene and N-Methylpyrrolidone, 1 mL each. Coloured extracts were centrifuged at 12’000 g, filtered with 0.45 µm PTFE syringe filters and analysed with Ion Pair Reversed Phase HPLC with Ultraviolet Diode Array (UV/DAD) detection under the following conditions: Kromasil-column C18, 5 µm, 150 × 2 mm (30°C); eluent A: aqueous solution of dodecyltrimethylammonium bromide (3 g/L) and ammonium bromide (1 g/L), eluent B: ethanolic solution of dodecyltrimethylammonium bromide (3 g/L) and ammonium bromide (1 g/L); run time = 30 minutes; flow rate = 0.35 mL/min; gradient conditions: 0 min 45% eluent B, 2 min 55% eluent B, 10 min 65% eluent B, 20 min 100% eluent B, 24 min 100% eluent B, 24.1 min 45% eluent B.

Results and discussion

Over the last eight years, more than 150 individual colour pigments and about 450 commercial tattoo ink or PMU samples from market surveys were subjected to LDI- or MALDI-TOF-MS analyses. Mass profiles were recorded either from pigments of different producers in order to build a home-made reference spectrum library or from samples taken during market surveys. From these former investigations we derived the following conclusions: LDI- or MALDI-TOF-MS is an excellent and rapid method for the detection and identification of organic colour pigments, but not for inorganic ones. For the most part of the analysed pigments, the ionisation can be performed without added matrix (LDI-TOF-MS), because the pigments themselves function as chromophores absorbing the laser beam14.

Method overview

After clean-up, samples were screened with LDI-TOF-MS using the NIST program with our pigment library for identification. Tentatively identified pigments were then confirmed by comparing the high-resolution mass spectra of the native and the reference spiked sample. As a further check, the resulting colour of found pigments had to give the sample colour. Whenever the results of the analysis did not match with the colour of the samples, the ionisation was optimised by changing the laser power and/or adding DHB as a matrix to the sample.

Effect of ethanolic sample clean-up

Modern tattoo inks not only contain pigments but also binders, solvents, surfactants, preservatives and thickening agents15. Surfactants adsorb to the surface of pigment agglomerates and decrease the surface tension of the solvent. This eliminates residing air bubbles and improves particle coating with other additives. Adsorbed surfactants, however, can interfere with analysis; spiking C.I. 74265 (pigment green 36) with the non-ionic tenside Triton X-100 leads to a nearly complete suppression of mass signals (Figure 1). This problem was met by washing samples with EtOH before LDI-TOF-MS, which proved to be effective in most cases16.

Figure 1.

Open in a new tab

Screening of samples with NIST software

With no commercial library of pigments available, we made our own library by exporting ASCII files of the mass spectra of analysed reference pigments to NIST compatible unit resolution mass peak lists. We preferred the NIST software for screening samples because its reversed match algorithm (R. Match) allows the simultaneous identification of several pigments in mixtures. This advantage is demonstrated for a red tattoo ink inFigure 2. Although the three existing pigments have overlapping ion clusters, all could be identified by their characteristic ions: C.I. 73915 (P.R. 122) with the ions [M+H]+ (m/z 341) and the sodium adduct [M+Na]+ (m/z 363), C.I. 21110 (P.O. 13) with the molecular ion [M+H]+ (m/z 623) and the two fragment ion clusters at m/z 437 and m/z 251 and C.I. 12475 (P.R. 170) with the ions [M+H]+ (m/z 455), [M+Na]+ (m/z 477) and the fragment ion m/z 318. As a further advantage of the NIST software compound specific information such as structural formula, molecular mass, synonyms, colour and CAS-no. can be added to the mass spectra library, and therefore also be used as search criterions. In the case of unknown mass spectra, the software can also extend the search to commercial GC-MS mass libraries (e.g. NIST main library).

Figure 2. Head to tail plot of three pigments identified in the mass spectra of a red tattoo ink sample by NIST software: C.I.73915 (Pigment Red 122, R. Match = 878), C.I.21110 (Pigment Orange 13, R. Match = 833) and C.I.12475 (Pigment Red 170, R. Match = 780).

Open in a new tab

Verification of pigments

Standard addition and high-resolution mass spectrometry were used for the verification of particularly interesting pigments tentatively identified by NIST. As a quality control standard addition also allows checking for possible discriminations caused by the sample matrix. Verification is shown for a violet coloured tattoo ink containing the prohibited C.I. 51319 (P.V. 23), which was not declared (Figure 3). After being tentatively identified in a first run, spiking the sample with C.I. 51319 (Figure 3B) showed the same typical molecular ion cluster [M]+ (m/z 588) and two fragment ion clusters with m/z 554 and m/z 520 as in the original sample (Figure 3A). Instead of C.I. 51319, the product label listed two legal pigments red C.I.12477 (P.R. 210) and blue C.I.74160 (P.B. 15) both non-present in the ink. LDI spectra did not reveal their characteristic ions [M+Na]+ (m/z 463) and [M+Na]+ (m/z 477) for the red pigment (a mixture of C.I. 12474 and 12475) and the ion cluster [M+H]+ (m/z 575) of the blue one (Figure 3C). This demonstrates a typical case of label forging for tattoo inks.

Figure 3.

Open in a new tab

Signal enhancement using DHB as matrix

In the case of some 3,3’-dichlorobenzidine based diazodiaryl-pigments, only weak molecular ion mass signals ([M]+, [M+H]+, [M+Na]+) but more pronounced fragmentation were observed. In order to enhance molecular mass signals, three of the most often used MALDI- matrices, dihydroxybenzoic acid (DHB), α-cyano-4-hydroxycinnamic acid (CCA), and sinapic acid (SA), were added to ink samples containing dichlorobenzidine based pigments. It turned out that all three had a positive effect, with DHB showing the best signal enhancement. The effect of DHB on the mass spectra of yellow pigment C.I. 21090 is shown inFigure 4. With DHB the signal intensity of the molecular ion [M]+ (m/z 628) was twice that of the sodium adduct [M+Na]+ (m/z 651), even twelve times as high as without matrix supplement.

Figure 4.

Open in a new tab

Method validation

Because the identification of pigments by LDI-TOF-MS is a qualitative method, validation was only performed for mass resolution, accuracy of pigment identification, a rough estimation of limits of detection and the precision of characteristic mass intensity.

At the mass of 450 amu, a mass resolution of R = M/ΔM = 6000 (full width at half maximum, FWHM) was achieved. The resolution quality depends on the laser power and is not the same for all pigments. Generally, mass resolution decreases with higher laser power and a shift to a higher mass of up to 0.1 u can be observed.

Accuracy of pigment identification in the screening mode (NIST, without sample washing and spiking) was tested with 48 different tattoo ink samples, their composition being unknown to the analyst. The results were checked for conformity using trustworthy declarations and/or additional in-house HPLC analysis of more or less soluble colourants. In 44 of the 48 samples (92%) all organic pigments were identified correctly (up to four in the same sample). Incomplete detection was found in two inks where the banned orange pigment C.I. 12075 remained undetected due to its weak signals in the mass spectra. The resulting poor library search matches did not give sufficient evidence for a positive identification. With the characteristic ion [M+H]+ (m/z 339), this pigment is usually well detectable. Therefore, a pigment concentration near the limit of detection is the most likely explanation for the weak signal. For the two other samples, results of LDI-TOF-MS were not consistent with those of HPLC-analysis: in the first case, C.I. 21110 was identified with LDI-TOF-MS, whereas HPLC-analysis gave C.I. 21115. In the second case C.I. 21100 was mistaken for C.I. 21095 because the characteristic mass signals of C.I. 21100 (m/z 522 and 550) were quenched.

The limits of detection (LOD) in a blue tattoo ink were estimated by standard addition with two red pigments, of which one was a well (C.I. 56110) and the other a poorly ionisable one (C.I. 14720). A correct identification of the pigment specific mass clusters by the NIST library was possible between sample concentrations of 1% (w/w) for the well ionisable pigment and of 10 – 20% (w/w) for the other one (C.I. 14720). As pigments are added to tattoo inks in the percent range, screening with LDI-TOF-MS proved to have a suitable LOD.

Nevertheless, we have to consider that sometimes the LOD can also be affected by discrimination effects of tensides or inorganic pigments, such as carbon black (C.I. 77266) or TiO2 (C.I. 77891, white), present in the sample16. Whereas tensides can be easily removed by EtOH the separation of the two inorganic pigments remain an unresolved problem up to this date. However, with the application of a higher laser energy and colour plausibility checks, negative identification results due to matrix effects could be eliminated for all forbidden organic pigments.

Precision was determined as repeatability with ten determinations of the characteristic mass intensity of two pigments (C.I. 21108, C.I. 74260) and two different inks containing C.I.74160 and C.I.12475, respectively. The pigment specific relative standard deviations were in a range of 4% to 13%, which are by far sufficient for identification purposes. This good reproducibility is in contrast to common experience with MALDI-TOF-MS as inhomogeneous crystallisation of the matrix promotes the building of so called “hot-spots” with locally higher ionisation signals17. Obviously, the dispersion of the pigments spotted on the target plate using MeOH instead of a matrix seems to give a more homogenous distribution of the analytes. Good reproducibility was probably also due to laser scanning in the serpentine style over the whole sample with 20 accumulated laser shots per profile which evens out some inhomogeneity.

Market survey

Over the last eight years we used LDI- or MALDI-TOF-MS for market survey purposes in Switzerland. Samples were randomly but also risk based collected from tattooing and PMU studios and from different importers.Table 1 gives an overview of the organic pigments identified in tattoo ink and PMU samples. Data obtained show that Pigment Blue 15 (C.I. 74160) was the most often found colourant followed by the prohibited Pigment Green 7 (C.I. 74260). All prohibited pigments, especially Pigment Green 7, were more frequently found in inks (3 – 13%) than in PMU (2 – 4%). In contrast to detected legal pigments where only 7% were missing on product labels of the samples, the prohibited pigments were often not declared (68%), for details see18. This clearly gives evidence that label forging is widespread. This fraud considers pigments that are banned due to toxicological concerns, showcasing the urgency for widespread market controls.

Table 1. Proportions of the most frequently identified pigments (≥ 2%) in about 450 products of Swiss market surveys between 2009 and 2017.

Prohibited colours are marked with an *.

PigmentNameTotal samples
(n=451)
Tattoo inks
(n=396)
Permanent make-up
(n=55)
C.I. 74160Pigment Blue 1527%28%16%
C.I. 74260*Pigment Green 712%13%4%
C.I. 12475Pigment Red 17010%11%4%
C.I. 56110Pigment Red 2549%8%16%
C.I. 73915*Pigment Red 1228%8%4%
C.I. 561170Pigment Orange 738%6%16%
C.I. 51319*Pigment Violet 238%8%2%
C.I. 11741Pigment Yellow 745%6%-
C.I. 21095Pigment Yellow 145%6%-
C.I. 21110Pigment Orange 135%6%-
C.I. 12477Pigment Red 2105%5%-
C.I. 56300Pigment Yellow 1384%5%-
C.I. 12490Pigment Red 54%2%24%
C.I. 74265Pigment Green 364%5%-
C.I. 51345Pigment Violet 373%4%-
C.I. 11767Pigment Yellow 973%3%4%
C.I. 73900*Pigment Violet 193%3%2%
C.I. 11740Pigment Yellow 653%3%-

Open in a new tab

MALDI raw data are provided as MASTER_RUN-, CAL-, COR-, LBL-, RAW, RUN, STATS- and UAP- files using Launchpad™ version 2.9.3 software (Shimadzu-Biotech Corp., Kyoto). The data include the raw high resolution spectra of Figures 1, 3 and 4

Copyright: © 2018 Niederer M et al.

Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).

Click here for additional data file. (542.6KB, tgz)

MALDI raw data are provided as TXT-files in ASCII-format using Launchpad™ version 2.9.3 software (Shimadzu-Biotech Corp., Kyoto). The data include the high resolution raw spectra of Figures 1, 3 and 4

Copyright: © 2018 Niederer M et al.

Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).

Click here for additional data file. (13.6KB, tgz)

NIST spectrum peak lists are provided as MSP-files in ASCII-format with a mass resolution of 1 D (msp) using the software of NIST 2.0 (USA). The data include the NIST spectra of Figure 2, Table 1 and method validation (HPLC verification)

Copyright: © 2018 Niederer M et al.

Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).

Click here for additional data file. (10.8KB, tgz)

Intermediary data of Table 1 and a list of suppliers of pigments of Table 1 are provided as Excel-files (XLS). The data include sample-nr, declaration of prohibited and legal pigments and reversed match values of pigments by NIST-software (Rfit)

Copyright: © 2018 Niederer M et al.

Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).

Click here for additional data file. (30.1KB, tgz)

An example of the construction of our home-made spectrum library using the NIST software 2.0 (USA) is shown as JPG-file. The data include compound name, structural formula, molecular mass, CAS-Nr., analysis date, intensity of the main spectrum peaks and synonyms

Copyright: © 2018 Niederer M et al.

Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).

Click here for additional data file. (112.9KB, tgz)

Conclusion

To our knowledge, no study reported a LDI-TOF-MS method for the identification of colour pigments in tattoo inks, PMU or cosmetic products. Method validation and using the LDI-TOF-MS method for market surveys demonstrated that the method described is fit for purpose. Therefore, LDI-TOF-MS is a further powerful tool particularly in combination with py-GC/MS10, HPLC or ultraviolet–visible spectroscopy18 for enforcing legal restrictions on pigments in tattoo inks and PMU. It should be kept in mind, however, that the presented method gives no information on the levels of the detected organic pigments. For this, further studies would be necessary.

Data availability

The data referenced by this article are under copyright with the following copyright statement: Copyright: © 2018 Niederer M et al.

Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).http://creativecommons.org/publicdomain/zero/1.0/

As a law enforcement body, we are bound to official secrecy. Scrutinising and anonymising data was only done within an acceptable effort. Considering these restrictions, we provide raw data for all experiments presented in our work (Figure 1Figure 4) and of the 18 most frequent pigments of the market surveys, as shown inTable 1. Further, we created an Excel sheet (Dataset 4) with intermediary data of the market surveys.

F1000Research: Dataset 1. MALDI raw data are provided as MASTER_RUN-, CAL-, COR-, LBL-, RAW, RUN, STATS- and UAP- files using Launchpad™ version 2.9.3 software (Shimadzu-Biotech Corp., Kyoto). The data include the raw high resolution spectra ofFigure 1,Figure 3 andFigure 4,10.5256/f1000research.13035.d18476619

F1000Research: Dataset 2. MALDI raw data are provided as TXT-files in ASCII-format using Launchpad™ version 2.9.3 software (Shimadzu-Biotech Corp., Kyoto). The data include the high resolution raw spectra ofFigure 1,Figure 3 andFigure 4,10.5256/f1000research.13035.d18476720

F1000Research: Dataset 3. NIST spectrum peak lists are provided as MSP-files in ASCII-format with a mass resolution of 1 D (msp) using the software of NIST 2.0 (USA). The data include the NIST spectra ofFigure 2,Table 1 and method validation (HPLC verification),10.5256/f1000research.13035.d18476821

F1000Research: Dataset 4. Intermediary data ofTable 1 and a list of suppliers of pigments ofTable 1 are provided as Excel-files (XLS). The data include sample-no, declaration of prohibited and legal pigments and reversed match values of pigments by NIST-software (Rfit),10.5256/f1000research.13035.d18476922

F1000Research: Dataset 5. An example of the construction of our home-made spectrum library using the NIST software 2.0 (USA) is shown as JPG-file. The data include compound name, structural formula, molecular mass, CAS-no., analysis date, intensity of the main spectrum peaks and synonyms,10.5256/f1000research.13035.d18477023

Acknowledgements

We thank Bernard Roux (State Laboratory Basel-City) for analytical support and helpful suggestions.

Funding Statement

The author(s) declared that no grants were involved in supporting this work.

[version 2; referees: 2 approved]

References

  • 1.Piccinini P, Contor L, Pakalin S, et al. :Safety of tattoos and permanent make-up. State of play and trends in tattoo practices. JRC technical report. EUR 27528 EN;2015. 10.2788/924128 [DOI] [Google Scholar]
  • 2.Council of Europe: Resolution ResAP(2008)1 on requirements and criteria for the safety of tattoos and permanent make-up. (Superseding resolution ResAP(2003)2 on tattoos and permanent make-up).Council Resolution of 20 February 2008. Accessed 15 Sep 2017.Reference Source [Google Scholar]
  • 3.Verordnung des EDI über Gegenstände für den Schleimhaut-, Haut- und Haarkontakt sowie über Kerzen, Streichhölzer, Feuerzeuge und Scherzartikel (817.023.41). Bern,2005. Accessed 15 Sep 2017.Reference Source [Google Scholar]
  • 4.Lehner K, Santarelli F, Penning R, et al. :The decrease of pigment concentration in red tattooed skin years after tattooing.J Eur Acad Dermatol Venereol.2011;25(11):1340–1345.10.1111/j.1468-3083.2011.03987.x [DOI] [PubMed] [Google Scholar]
  • 5.Schreiver I, Hesse B, Seim C, et al. :Synchrotron-based ν-XRF mapping and μ-FTIR microscopy enable to look into the fate and effects of tattoo pigments in human skin.Sci Rep.2017;7(1):11395.10.1038/s41598-017-11721-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Hauri U, Hohl C:Photostability and breakdown products of pigments currently used in tattoo inks.Curr Probl Dermatol.2015;48:164–169.10.1159/000369225 [DOI] [PubMed] [Google Scholar]
  • 7.Poon KWC, Dadour IR, McKinley AJ:In situ chemical analysis of modern organic tattooing inks and pigments by micro-Raman spectroscopy.J Raman Spectrosc.2008;39(9):1227–1237.10.1002/jrs.1973 [DOI] [Google Scholar]
  • 8.Vila A, Garcia JF:Analysis of the chemical composition of red pigments and inks for the characterization and differentiation of contemporary prints.Analytical Letters.2012;45(10):1274–1285.10.1080/00032719.2012.673100 [DOI] [Google Scholar]
  • 9.Hutton Carlsen K, Køcks M, Sepehri M, et al. :Allergic reactions in red tattoos: Raman spectroscopy for 'fingerprint' detection of chemical risk spectra in tattooed skin and culprit tattoo inks.Skin Res Technol.2016;22(4):460–469.10.1111/srt.12287 [DOI] [PubMed] [Google Scholar]
  • 10.Schreiver I, Hutzler C, Andree S, et al. :Identification and hazard prediction of tattoo pigments by means of pyrolysis-gas chromatography/mass spectrometry.Arch Toxicol.2016;90(7):1639–50.10.1007/s00204-016-1739-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Stachura S, Desiderio VJ, Allison J:Identification of organic pigments in automotive coatings using laser desorption mass spectrometry.J Forensic Sci.2007;52(3):595–603.10.1111/j.1556-4029.2007.00431.x [DOI] [PubMed] [Google Scholar]
  • 12.Soltzberg LJ, Hagar A, Kridaratikon S, et al. :MALDI-TOF mass spectrometric identification of dyes and pigments.J Am Soc Mass Spectrom.2007;18(11):2001–2006.10.1016/j.jasms.2007.08.008 [DOI] [PubMed] [Google Scholar]
  • 13.Boon JJ, Learner T:Analytical mass spectrometry of artists’ acrylic emulsion paints by direct temperature resolved mass spectrometry and laser desorption ionisation mass spectrometry.J Anal Appl Pyrolysis.2002;64(2):327–344.10.1016/S0165-2370(02)00045-1 [DOI] [Google Scholar]
  • 14.Wyplosz N:Laser desorption mass spectrometric studies of artists' organic pigments.PhD Thesis University of Amsterdam;2003. Reference Source [Google Scholar]
  • 15.Dirks M:Making innovative tattoo ink products with improved safety: possible and impossible ingredients in practical usage.In: Serup J, Kluger N, Bäumler W, editors.Tattooed Skin and Health. Curr Probl Dermatol.Basel: Karger AG;2015;48:118–127.10.1159/000369236 [DOI] [PubMed] [Google Scholar]
  • 16.Lechner L:Analyse von Pigmenten in Tätowiertinten mit MALDI-TOF-MS. Wissenschaftliche Abschlussarbeit Universität Hohenheim Stuttgart;2015. [Google Scholar]
  • 17.Persike M:Verfahrensentwicklung zur Analytik kleiner Moleküle mittels MALDI-Massenspektrometrie. Dissertation Johann Wolfgang Goethe-Universität Frankfurt am Main;2010. Reference Source [Google Scholar]
  • 18.Hauri U:Inks for tattoos and permanent make-up—pigments, preservatives, aromatic amines, polyaromatic hydrocarbons and nitrosamines.Swiss National Investigation Campaign 2014. Department of Health, Kanton Basel-Stadt.2014. [Google Scholar]
  • 19.Niederer M, Hauri U, Kroll L, et al. :Dataset 1 in: Identification of organic pigments in tattoo inks and permanent make-up using laser desorption ionisation mass spectrometry.F1000Research.2017. Data Source [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Niederer M, Hauri U, Kroll L, et al. :Dataset 2 in: Identification of organic pigments in tattoo inks and permanent make-up using laser desorption ionisation mass spectrometry.F1000Research.2017. Data Source [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Niederer M, Hauri U, Kroll L, et al. :Dataset 3 in: Identification of organic pigments in tattoo inks and permanent make-up using laser desorption ionisation mass spectrometry.F1000Research.2017. Data Source [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Niederer M, Hauri U, Kroll L, et al. :Dataset 4 in: Identification of organic pigments in tattoo inks and permanent make-up using laser desorption ionisation mass spectrometry.F1000Research.2017. Data Source [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Niederer M, Hauri U, Kroll L, et al. :Dataset 5 in: Identification of organic pigments in tattoo and permanent make-up inks using laser desorption ionisation mass spectrometry.F1000Research.2017. Data Source [DOI] [PMC free article] [PubMed] [Google Scholar]

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    MALDI raw data are provided as MASTER_RUN-, CAL-, COR-, LBL-, RAW, RUN, STATS- and UAP- files using Launchpad™ version 2.9.3 software (Shimadzu-Biotech Corp., Kyoto). The data include the raw high resolution spectra of Figures 1, 3 and 4

    Copyright: © 2018 Niederer M et al.

    Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).

    Click here for additional data file. (542.6KB, tgz)

    MALDI raw data are provided as TXT-files in ASCII-format using Launchpad™ version 2.9.3 software (Shimadzu-Biotech Corp., Kyoto). The data include the high resolution raw spectra of Figures 1, 3 and 4

    Copyright: © 2018 Niederer M et al.

    Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).

    Click here for additional data file. (13.6KB, tgz)

    NIST spectrum peak lists are provided as MSP-files in ASCII-format with a mass resolution of 1 D (msp) using the software of NIST 2.0 (USA). The data include the NIST spectra of Figure 2, Table 1 and method validation (HPLC verification)

    Copyright: © 2018 Niederer M et al.

    Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).

    Click here for additional data file. (10.8KB, tgz)

    Intermediary data of Table 1 and a list of suppliers of pigments of Table 1 are provided as Excel-files (XLS). The data include sample-nr, declaration of prohibited and legal pigments and reversed match values of pigments by NIST-software (Rfit)

    Copyright: © 2018 Niederer M et al.

    Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).

    Click here for additional data file. (30.1KB, tgz)

    An example of the construction of our home-made spectrum library using the NIST software 2.0 (USA) is shown as JPG-file. The data include compound name, structural formula, molecular mass, CAS-Nr., analysis date, intensity of the main spectrum peaks and synonyms

    Copyright: © 2018 Niederer M et al.

    Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).

    Click here for additional data file. (112.9KB, tgz)

    Data Availability Statement

    The data referenced by this article are under copyright with the following copyright statement: Copyright: © 2018 Niederer M et al.

    Data associated with the article are available under the terms of the Creative Commons Zero "No rights reserved" data waiver (CC0 1.0 Public domain dedication).http://creativecommons.org/publicdomain/zero/1.0/

    As a law enforcement body, we are bound to official secrecy. Scrutinising and anonymising data was only done within an acceptable effort. Considering these restrictions, we provide raw data for all experiments presented in our work (Figure 1Figure 4) and of the 18 most frequent pigments of the market surveys, as shown inTable 1. Further, we created an Excel sheet (Dataset 4) with intermediary data of the market surveys.

    F1000Research: Dataset 1. MALDI raw data are provided as MASTER_RUN-, CAL-, COR-, LBL-, RAW, RUN, STATS- and UAP- files using Launchpad™ version 2.9.3 software (Shimadzu-Biotech Corp., Kyoto). The data include the raw high resolution spectra ofFigure 1,Figure 3 andFigure 4,10.5256/f1000research.13035.d18476619

    F1000Research: Dataset 2. MALDI raw data are provided as TXT-files in ASCII-format using Launchpad™ version 2.9.3 software (Shimadzu-Biotech Corp., Kyoto). The data include the high resolution raw spectra ofFigure 1,Figure 3 andFigure 4,10.5256/f1000research.13035.d18476720

    F1000Research: Dataset 3. NIST spectrum peak lists are provided as MSP-files in ASCII-format with a mass resolution of 1 D (msp) using the software of NIST 2.0 (USA). The data include the NIST spectra ofFigure 2,Table 1 and method validation (HPLC verification),10.5256/f1000research.13035.d18476821

    F1000Research: Dataset 4. Intermediary data ofTable 1 and a list of suppliers of pigments ofTable 1 are provided as Excel-files (XLS). The data include sample-no, declaration of prohibited and legal pigments and reversed match values of pigments by NIST-software (Rfit),10.5256/f1000research.13035.d18476922

    F1000Research: Dataset 5. An example of the construction of our home-made spectrum library using the NIST software 2.0 (USA) is shown as JPG-file. The data include compound name, structural formula, molecular mass, CAS-no., analysis date, intensity of the main spectrum peaks and synonyms,10.5256/f1000research.13035.d18477023

    Identification of organic pigments in tattoo inks and permanent make-up using laser desorption ionisation mass spectrometry (2024)

    References

    Top Articles
    Latest Posts
    Recommended Articles
    Article information

    Author: Fr. Dewey Fisher

    Last Updated:

    Views: 6624

    Rating: 4.1 / 5 (42 voted)

    Reviews: 81% of readers found this page helpful

    Author information

    Name: Fr. Dewey Fisher

    Birthday: 1993-03-26

    Address: 917 Hyun Views, Rogahnmouth, KY 91013-8827

    Phone: +5938540192553

    Job: Administration Developer

    Hobby: Embroidery, Horseback riding, Juggling, Urban exploration, Skiing, Cycling, Handball

    Introduction: My name is Fr. Dewey Fisher, I am a powerful, open, faithful, combative, spotless, faithful, fair person who loves writing and wants to share my knowledge and understanding with you.